Hot sales Fuji ZRM/ZKM Zirconia oxygen analyzer
100% New & Original with competitive price and fast delivery,one year warranty!
Zirconia oxygen analyzer is used to continuously measure the oxygen content in the combustion exhaust gas of industrial boilers, heating furnaces, cracking furnaces, incinerators, etc., through monitoring and control of the combustion process, so as to achieve energy saving and high efficiency, reduce environmental pollution and prolong furnace age purpose.
The zirconia analyzers we provide mainly include four series of products: ordinary type, flameproof type, high dust-resistant type and two-component zirconia analyzer according to different application occasions. We also provide customers with functional safety zirconia analyzers. , improve the reliability of measurement and control systems.
Zirconia can also be used for process gas measurement of non-combustion gas to monitor and control the oxygen content (such as tight transport of catalysts, nitrogen sealing of metal magnesium), and we also have a pressure compensation calibration box to ensure accuracy of positive pressure measurement.
Detector Specifications (ZFK)
Measuring gas temperature: guide tube type: - 10 ~ + 600 ℃ (for general use, for corrosion resistance)
Ejector type: -10 to +1500℃ (for high temperature)
- 10 to + 800℃ (for general use)
Measuring gas pressure: - 3 ~ + 3kPa
Guide tube without blow-out nozzle, with blow-out nozzle
Flange: JIS5K 65A FF
Insertion length: 0.3, 0.5, 0.75, 1m
For high dust (with blow-out nozzle)
Without cover, with cover
Flange: JIS5K 80A FF
Insertion length: 0.8m (standard)
Injector: The probe that introduces the measuring gas into the detector
Flange: JIS10K 65A RF
Insertion length: 0.5, 0.75, 1, 1.5m (can)
Operating temperature: detection part: -10 ~ +60 ℃,
Ejector section: -5 to +100°C
Detector flange surface: 125°C or less when energized
Storage temperature: Detection section: -20 to +70°C
Ejector part: -10 to +100℃
Structure: Dust-proof, rain-proof structure (equivalent to IEC standard IP66, however, the filter part of the outlet head) In cold areas, a thermal cover is required (can be)
Filter: Alumina (filtration accuracy 50μm) and quartz filter paper
The main material of the gas connection part: Detector: Zirconium oxide, SCS14 (equivalent to SUS316), platinum, SUS304
Injector (general use): SUS316 , SUS304
Injector (for high temperature): SiC, SUS316 SUS304
Calibration gas port: φ 6mm hose connector or φ 1/4 inch hose connector (optional)
Standard gas port: φ 6mm hose fitting or ø 1/4 inch hose fitting (available) Use when the reference oxygen concentration changes, etc.
Detector installation: horizontal ±45°, clean air around.
Dimensions: (L × maximum diameter) 210mm × 100mm
Mass: detector about 1.6kg
Injector Approx. 15kg (insertion length 1m)
Guide tube (for general use) 1m, about 5kg
Spray color: silver gray and SUS metallic color
Ejector air inlet: (flow) 5 ~ 10L/min
Calibration gas flow: 1.5 ~ 2L/min
Blowing air inlet: 200 ~ 300kPa
Ejector exhaust gas treatment: circulating in furnace and flue
Heater temperature drop alarm output (injector): Output alarm when +100°C or less Mechanical thermostat 1a Contact AC 200V 2A
Inverter Specifications (Model: ZKM) ・
Concentration value display: digital 4-digit display ・
Contact output signal: (1) Contact specification 6 points 1a AC 250V/3A or DC30V/3A
(2) Contact function ・ under maintenance
・Instrument abnormality Note 1)
・Alarm Note 2)
・ Zero calibration gas
・ Full-scale calibration gas
・Blowing out Note 3)
Note 1) The instrument abnormality is 1) thermocouple disconnection, 2) sensor disconnection, 3) temperature abnormality, 4) calibration abnormality, 5) zero point and full scale abnormality, and 6) the contact is ON in case of output abnormality
Note 2) Alarm selection 1) upper limit, 2) lower limit, 3) upper and lower limit, 4) upper and lower alarms, 5) any one point of lower and lower alarms, ON during operation
Note 3) It is valid in the optional case during blowing, and ON during operation
Contact input signal:
(1) Contact specification 3 points (optionally selected according to the following conditions)
ON; 0V, OFF; 5V (below 10mA when ON)
(2) Contact function ・ External hold ・ Calculation reset ・ Heater OFF ・ Blow off (optional) ・ Calibration inhibit ・ Calibration start ・ Range switching
Calibration method: (a) Manual calibration through keyboard operation
(b) Automatic calibration (optional) Calibration cycle: 00 days 00 hours to 99 days 23 hours
(c) Correction together
Calibration gas: The range that can be set Zero gas 0.010 ~ 25.00%O2 Full scale gas 0.010 ~ 50.00%O2 *Calibration gas concentration Zero gas 0.25 ~ 2.0%O2 Full scale gas 20.6 ~ 21.0%O2 (oxygen concentration in the atmosphere)
Blow out: This function can use compressed air to concentrate the guide tube
(Optional) The collected dust is blown away.
It can be blown out according to the time interval set in advance and according to the set time.
Blowing cycle: 00 hours 00 minutes to 99 hours 59 minutes
Blowing time: 0 seconds to 999 seconds
Output hold: manual calibration, automatic calibration, blowing, transmission
Output hold, hold release, etc. can be performed during sensor reactivation and warm-up
Switching valve: During manual zero-point and full-scale calibration, the switching valve can
(Optional) To switch between zero and span gas. Mounted on the side of the converter.
Communication function: RS232C (MODBUS) standard specification RS-485 (MODBUS) (optional) ・
Combustion Efficiency Display: When this display is selected, the rich mode will be displayed at the same time. (Optional) This function can calculate and display the combustion efficiency from the oxygen concentration value and the measured gas temperature. A separate R thermocouple is required for temperature measurement. ・Operating temperature: -20 to +55℃ ・Operating humidity: 95%RH or less without condensation ・Storage temperature: -30~+70℃ ・Storage humidity: 95%RH or less without condensation
Structure: Dust- and rain-proof structure (equivalent to IEC standard IP66 or IP67)
Material: aluminum shell
Dimensions (H × W × D): 170 × 159 × 70mm (IP66) 220 × 230 × 95mm (IP67)
Mass: about 2kg (excluding cable, detector) (IP66)
Approx. 4.5kg (excluding cable, detector) (IP67)
Paint Color: Housing: Silver Gray Cover: Pantone Color Code 1C-F (Cool Gray) (Small Housing) Marshall 6PB 3.5/10.5 (Blue) (Large Housing)
Installation method: panel surface installation, or pipe installation
Corresponding model Typical model
Zirconia detector ZFK8R ZFK8K135-0Y0YY-1C2
Inverter ZKM ZKMABY12-YCY1A-YRY
Cable ZRZKR ZRZKRYB1-2
Injector ZTA ZTA
Zirconia detector ZFK8R ZFK8K135-0Y0YY-1C2
Inverter ZKM ZKM1BY11-YCY1
Cable ZRZKR ZRZKRYBA-1
Injector ZTA ZTA
Explosion-proof detector ZFKER
Explosion-proof converter ZKME
Explosion-proof cable ZRZER
Zirconia Oxygen Analyzer Converter ZRM
fuse tube
Ceramic filter ZZPZFK3-A090
Detector O-ring ZZPZFK3-A100
Detector ZFK□R□34-0Y0YY
duct
Check box (manual automatic double loop) YXX
Explosion-proof zirconia analyzer ZRM-B ZRY-B
Zirconia Converter ZRM
Zirconia Converter ZRY
Detector ZFK□□□□4-□□□□□-□
Replacement detector maintenance ZFK□□□□4-0Y0YY
Converter ZRM1□□□1-□□
Injector ZTA□1□□1
Extension cable ZRZ□□□□1-□
Calibration box YXX□1-□□
Converter ZRY1BY12-1R□□
Detector ZFK□□□□5-0Y0YY-□□□
Guide tube ZTB□□-□□□□-□□□
Extension cable ZRZ□□□□1-□
Converter ZRM1□□□1-3□
Injector ZTA□1-□□□□ □□□
Detection box ZMZ□□□□□